Proton transfer reaction time-of-flight mass spectrometry (PTR-TOF-MS) allows for very fast simultaneous monitoring of volatile organic compounds (VOCs) in complex environments. In several applications, food science and food technology in particular, peaks with very different intensities are present in a single spectrum. For VOCs, the concentrations range from the sub-ppt all the way up to the ppm level. Thus, a large dynamic range is necessary. In particular, high intensity peaks are a problem because for them the linear dependency of the detector signal on VOC concentration is distorted. In this paper we present, test with real data, and discuss a novel method which extends the linearity of PTR-TOF-MS for high intensity peaks far beyond the limit allowed by the usual analytical correction methods such as the so-called Poisson correction. Usually, raw data can be used directly without corrections with an intensity of up to about 0.1 ions/pulse, and the Poisson correction allows the use of peaks with intensities of a few ions/pulse. Our method further extends the linear range by at least one order of magnitude. Although this work originated from the necessity to extend the dynamic range of PTR-TOF-MS instruments in agro-industrial applications, it is by no means limited to this area, and can be implemented wherever dead time corrections are an issue.

Cappellin, L.; Biasioli, F.; Schuhfried, E.; Soukoulis, C.; Märk, T.D.; Gasperi, F. (2011). Extending the dynamic range of proton transfer reaction time-of-flight mass spectrometers by a novel dead time correction. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 25 (1): 179-183. doi: 10.1002/RCM.4819 handle: http://hdl.handle.net/10449/19757

Extending the dynamic range of proton transfer reaction time-of-flight mass spectrometers by a novel dead time correction

Cappellin, Luca;Biasioli, Franco;Schuhfried, Erna;Soukoulis, Christos;Gasperi, Flavia
2011-01-01

Abstract

Proton transfer reaction time-of-flight mass spectrometry (PTR-TOF-MS) allows for very fast simultaneous monitoring of volatile organic compounds (VOCs) in complex environments. In several applications, food science and food technology in particular, peaks with very different intensities are present in a single spectrum. For VOCs, the concentrations range from the sub-ppt all the way up to the ppm level. Thus, a large dynamic range is necessary. In particular, high intensity peaks are a problem because for them the linear dependency of the detector signal on VOC concentration is distorted. In this paper we present, test with real data, and discuss a novel method which extends the linearity of PTR-TOF-MS for high intensity peaks far beyond the limit allowed by the usual analytical correction methods such as the so-called Poisson correction. Usually, raw data can be used directly without corrections with an intensity of up to about 0.1 ions/pulse, and the Poisson correction allows the use of peaks with intensities of a few ions/pulse. Our method further extends the linear range by at least one order of magnitude. Although this work originated from the necessity to extend the dynamic range of PTR-TOF-MS instruments in agro-industrial applications, it is by no means limited to this area, and can be implemented wherever dead time corrections are an issue.
2011
Cappellin, L.; Biasioli, F.; Schuhfried, E.; Soukoulis, C.; Märk, T.D.; Gasperi, F. (2011). Extending the dynamic range of proton transfer reaction time-of-flight mass spectrometers by a novel dead time correction. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 25 (1): 179-183. doi: 10.1002/RCM.4819 handle: http://hdl.handle.net/10449/19757
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10449/19757
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 62
  • ???jsp.display-item.citation.isi??? 56
social impact